Abstract:
The relationship between electrode dislocation and operative circumstances and case details was investigated in 486 patients who had pacemakers implanted. Electrode dislocation was more common among women and unrelated to age. The greater the operator's experience the lower the dislocation rate. Best results were obtained when implantation was performed via the left cephalic vein. There was a close relationship between threshold measurement obtained at the time of electrode placement and the frequency of dislocation.