Initial experience with oral contrast in PET/CT: phantom and clinical studies
Christian Cohade1, Medhat Osman, Yuji Nakamoto
1Division of Nuclear Medicine, Russell H. Morgan Department of Radiology and Radiological Sciences, Johns Hopkins Hospital, Baltimore, Maryland 21287-0817, USA.
Unlabelled:
The aims of the study were to evaluate the effects of oral contrast on apparent tracer activity measured with PET/CT when using CT attenuation correction and to report our initial experience in the use of oral contrast with PET/CT.
Methods:
Phantom studies with (18)F activity and saline bags or syringes filled with barium or gastrografin of varying densities were performed using a PET/CT scanner (CT attenuation correction). In the study, 91 clinical patients received dilute oral contrast and were evaluated by whole-body (18)F-FDG PET.
Results:
A phantom experiment with CT contrast (1.3% weight/volume [w/v] barium) showed a "cold" area in the cold stomach whereas a phantom with high-density barium (98% w/v) showed an artifactual focus of intense "activity" in the cold stomach. In clinical studies, stomach and right colon were opacified by CT contrast. Maximal measured contrast density was 239 Hounsfield units.
Conclusion:
High-density barium causes overestimation of tissue (18)F-FDG concentration. Low-density barium does not cause significant artifacts and appears suitable for clinical use.