Related Experiment Video
Updated: Aug 14, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Unusual extraction of an active fixation ventricular pacing lead with outer coil fracture in a child
T Karagoz1, A Celiker, O Hallioglu
1Paediatric Cardiology Unit, Hacettepe University Faculty of Medicine, Sihhiye, Ankara 06100, Turkey.
Aim:
In this report we describe an asymptomatic paced child with outer coil fracture of the bipolar screw-in steroid eluting ventricular lead without insulation defect, loss of ventricular capture and unexpected increase in lead impedance in the bipolar VVIR pace configuration.
Method And Result:
A previously unpublished method was performed to retract the helix and the lead. As a result, the fractured lead was successfully retrieved.
Conclusion:
We suggest that, this unusual extraction method can be tried as an alternative approach in the removal of an active fixation ventricular pacing lead with impaired mechanical function possibly due to fracture.