Compensating for the effect of inlet gas temperature on heated humidifier performance
Related Experiment Video
Updated: Sep 27, 2026

Temperature-Controlled Assembly and Characterization of a Droplet Interface Bilayer
Published on: April 19, 2021
1Neonatal and Paediatric Intensive Care Unit, Royal Childrens Hospital, Melbourne, Victoria.
Abstract:
The humidity output of heated humidifiers may be compromised by inlet gas temperatures exceeding approximately 26 degrees C, with humidity dropping below the recommended levels for intubated patients. A new version of the Fisher & Paykel MR850 humidifier claims to deal with this problem by offering a humidity compensation option. The present study tested this feature by measuring humidity output using the gravimetric method and a hygrometer at different inlet gas temperatures (16.6 degrees C to 40.0 degrees C) with compensation on and off. It was found that the compensation is effective in maintaining humidity levels despite high inlet gas temperatures.
Related Concept Videos
Joule-Thomson Effect
This experiment forces high-pressure gas through a throttle valve or a porous plug to a lower-pressure region. The gas expands as it passes through to...
The Joule and Joule–Thomson Experiments
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Efficiency of The Carnot Cycle
Heat Capacities of an Ideal Gas I
Molar heat capacity quantifies the ratio of the amount of heat added (or removed) to increase (or decrease) the temperature of...
Heat Capacities of an Ideal Gas II