Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
Near-patient testing for coagulopathy after cardiac surgery
R R Johi1, M H Cross, S D Hansbro
1Department of Anaesthesia, Leeds General Infirmary, Great George Street, Leeds LS1 3EX, UK.
The Hemochron Response system showed good agreement for prothrombin time (PT) and acceptable activated partial thromboplastin time (APTT) before cardiopulmonary bypass. However, accuracy decreased for both PT and APTT after bypass.
Area of Science:
- Clinical chemistry
- Medical device evaluation
- Cardiovascular surgery
Background:
- Near-patient testing systems aim to provide rapid diagnostic results at the point of care.
- Accurate coagulation monitoring is crucial for patients undergoing cardiac surgery with cardiopulmonary bypass.
Purpose of the Study:
- To evaluate the accuracy and precision of the Hemochron Response near-patient testing system.
- To compare Hemochron Response measurements of prothrombin time (PT) and activated partial thromboplastin time (APTT) with laboratory assays.
Main Methods:
- Fifty patients undergoing cardiac surgery with cardiopulmonary bypass were studied.
- Prothrombin time (PT) and activated partial thromboplastin time (APTT) were measured using the Hemochron Response and laboratory assays.
- Measurements were taken before and after cardiopulmonary bypass, including after protamine administration.
Main Results:
- Before bypass, PT measurements showed good agreement (bias +0.34) with small limits of agreement.
- After bypass, PT measurements exhibited increased bias (-3.27) and wider limits of agreement.
- Activated partial thromboplastin time (APTT) measurements demonstrated wider bias and limits of agreement compared to PT, both before and after bypass.
Conclusions:
- The Hemochron Response system demonstrated good agreement for PT and clinically acceptable agreement for APTT pre-bypass.
- Post-bypass, both PT and APTT measurements showed increased bias and limits of agreement, potentially becoming clinically unacceptable.
- The accuracy of the Hemochron Response system may be compromised in the post-cardiopulmonary bypass setting.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Cardiomyopathy VII: Pre and Post Operative Nursing Management