Related Experiment Video
Updated: Sep 26, 2026

Electrophysiological Methods to Assess Peripheral Pain Block in an Anesthetized Rat
Published on: November 21, 2025
Nerve stimulators used for peripheral nerve blocks vary in their electrical characteristics
Admir Hadzic1, Jerry Vloka, Nihad Hadzic
1Department of Anesthesiology, St. Luke's-Roosevelt Hospital Center, New York, NY 10025, USA. ah149@columbia.edu
Background:
Nerve stimulation with a low-intensity electrical current has become a vital part of the performance of peripheral nerve blockade. The purpose of this study was to compare the accuracy and characteristics of peripheral nerve stimulators used in clinical practice in the United States.
Methods:
Fifteen peripheral nerve stimulators were fitted with fresh batteries and set to deliver currents ranging from 0.1 to 4.0 mA into a series of high-tolerance resistance loads ranging from 1 to 100 komega. The current output, stimulus duration, morphology, frequency, and maximum voltage output were studied using a factory-calibrated oscilloscope.
Results:
All peripheral nerve stimulators performed uniformly well when set to deliver currents of 1.0 mA or more into a standard resistance load of 1 or 2 komega. However, at lower currents, the median error (%) increased from 2.4 (-5-144%) at 0.5 mA to 10.4 (-24-180%) at 0.1 mA into a 1 komega load. The morphology of the stimulus was characterized by a regular monophasic square pulse at current outputs of up to 1 mA and at a resistance of 1 komega. The stimulus waveform became particularly distorted as the impedance load was increased. The duration of the default stimulus set by the manufacturer varied from 34.8 to 460 micros among the peripheral nerve stimulators tested. The maximum voltage output ranged from 7.4 to 336 Volts.
Conclusions:
Nerve stimulators used for regional anesthesia vary greatly in accuracy of current output and in manufacturer-selected electrical characteristics (e.g., current duration, stimulating frequency, maximum voltage output).
Related Concept Videos
Local Anesthetics: Differential Sensitivity of Nerve Fibers
Nondepolarizing (Competitive) Neuromuscular Blockers: Pharmacological Actions
Although all competitive neuromuscular blockers are designed...
Nondepolarizing (Competitive) Neuromuscular Blockers: Mechanism of Action
Competitive antagonists prevent acetylcholine from binding to its receptor, inhibiting membrane depolarization. Without conformational changes or intrinsic...
Depolarizing Blockers: Mechanism of Action
Succinylcholine is the most commonly used depolarizing blocker. Chemically, it constitutes two molecules of acetylcholine joined together by an acetate methyl group. They act on the receptors in the same way as acetylcholine. Because succinylcholine...
Depolarizing Blockers: Pharmocokinetics
Local Anesthetics: Clinical Application as Surface, Infiltration, and Conduction Block Anesthesia