Related Experiment Videos
Performance evaluation of two multiplexes used in fluorescent short tandem repeat DNA analysis
Edward L Buse1, Jeanne C Putinier, Mary M Hong
1Forensic Science Services, DNA Section, Orange County Sheriff-Coroner Department, 320 N. Flower Street, Santa Ana, CA 92703, USA. elb@fss.co.orange.ca.us
Journal of Forensic Sciences
|April 1, 2003
Summary
This study validates two multiplex kits for short tandem repeat (STR) DNA profiling, essential for forensic labs and the CODIS database. The results confirm reliable DNA analysis using capillary electrophoresis (CE).
Area of Science:
- Forensic Science
- Molecular Biology
- Genetics
Background:
- Forensic laboratories rely on accurate short tandem repeat (STR) DNA profiling for criminal investigations.
- The U.S. national Combined DNA Indexing System (CODIS) utilizes 13 core STR loci for DNA identification.
- Validation of commercial DNA typing kits is crucial for ensuring reliable forensic analysis.
Purpose of the Study:
- To evaluate the performance of two commercial multiplex STR kits: AmpFlSTR Profiler Plus and AmpFlSTR COfiler.
- To validate the use of capillary electrophoresis (CE) with an ABI Prism 310 Genetic Analyzer for forensic DNA analysis.
- To assess key performance parameters including sensitivity, resolution, precision, and reliability with forensic samples.
Main Methods:
- Evaluation of AmpFlSTR Profiler Plus and COfiler kits for amplifying 13 core STR loci.
- Analysis of PCR products using capillary electrophoresis (CE) on an ABI Prism 310 Genetic Analyzer.
- On-site validation including sensitivity, resolution, binning, peak height ratios, mixtures, stutter, and environmental sample testing.
Main Results:
- Both multiplex kits demonstrated robust amplification and specificity to human DNA.
- Capillary electrophoresis consistently resolved fragments differing by one base, with acceptable binning by Genotyper software.
- Accurate typing results were achieved for non-probative, simulated, and environmentally challenged forensic samples.
Conclusions:
- AmpFlSTR Profiler Plus and COfiler kits are reliable for analyzing forensic DNA samples using CE.
- The validated methods support consistent and accurate STR profiling in routine forensic casework.
- These findings affirm the suitability of the evaluated kits and CE technology for forensic DNA databases like CODIS.