Evaluation of a new spirometer based on a fluidistor technique.

G Hedenstierna, S Lundberg, D Rawlings

    Acta Anaesthesiologica Scandinavica
    |January 1, 1976
    PubMed
    Summary

    A new fluidistor spirometer shows accuracy within +/-5% for normal breathing volumes compared to existing devices. Modifications could improve its performance at low and high flow rates for clinical use.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Comparison of the timing of development of immunity in Romney lambs from resistant and resilient selection lines.

    Veterinary parasitology·2019
    Same author

    Effect of Intraoperative High Positive End-Expiratory Pressure (PEEP) With Recruitment Maneuvers vs Low PEEP on Postoperative Pulmonary Complications in Obese Patients: A Randomized Clinical Trial.

    JAMA·2019
    Same author

    Tidal changes in PaO<sub>2</sub> and their relationship to cyclical lung recruitment/derecruitment in a porcine lung injury model.

    British journal of anaesthesia·2019
    Same author

    Specific anesthesia-induced lung volume changes from induction to emergence: a pilot study.

    Acta anaesthesiologica Scandinavica·2017
    Same author

    Atelectasis is inversely proportional to transpulmonary pressure during weaning from ventilator support in a large animal model.

    Acta anaesthesiologica Scandinavica·2017
    Same author

    Erratum to Protective intraoperative ventilation with higher versus lower levels of positive end-expiratory pressure in obese patients (PROBESE): study protocol for a randomized controlled trial.

    Trials·2017

    Area of Science:

    • Medical Devices
    • Respiratory Physiology

    Background:

    • Spirometry is crucial for assessing lung function.
    • Existing spirometers have limitations in accuracy and usability.

    Purpose of the Study:

    • To evaluate a novel spirometer based on fluidistor principles.
    • To assess its accuracy and suitability for clinical applications.

    Main Methods:

    • Evaluation using a mechanical model and human subjects.
    • Comparison with a Bernstein spirometer at various minute volumes.
    • Analysis of factors influencing measurement error.

    Main Results:

    • Accurate within +/-5% at normal minute volumes.
    • Underread at low volumes and overread at high volumes.

    Related Experiment Videos

  • Error increased with prolonged expiratory time and low-density gases.
  • Conclusions:

    • The fluidistor spirometer is stable, easy to handle, and accurate enough for anesthesia and intensive care.