Comparing the methods of measuring compartment pressures in acute compartment syndrome
Related Experiment Video
Updated: Sep 26, 2026

A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Annemarie Uliasz1, Jay T Ishida, Jason K Fleming
1Department of Pediatrics, University of Hawaii John A. Burns School of Medicine, Kapiolani Medical Center for Women and Children, Honolulu, Hawaii 96826, USA.
Abstract:
The objectives of this study were to compare the reliability and accuracy of published methods of measuring intramuscular compartment pressures. A muscle compartment model was constructed. Pressure was standardized using a water column exerting pressure on a muscle slab. The following three methods of intracompartmental pressure were used to measure the pressure within the muscle slab of the model: (1) Stryker, (2) manometric IV pump, and (3) Whitesides method. The Stryker and IV pump methods were similarly accurate compared with the "standard pressure." The Whitesides method was not reliable in measuring a reproducible pressure. The IV pump method is comparable to the Stryker instrument in terms of accuracy and ease of use. If a Stryker instrument is unavailable, the IV pump method can serve as a suitable and satisfactory alternative.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Measurement of Blood Pressure