Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Evaluation of a small cadmium zinc telluride detector for scintimammography
Bryon Mueller1, Michael K O'Connor, Ira Blevis
1Section of Nuclear Medicine, Department of Radiology, Mayo Clinic, Rochester, MN 55905, USA.
Unlabelled:
The purpose of this study was to evaluate a small semiconductor-based gamma camera that may have applications in scintimammography.
Methods:
A small cadmium zinc telluride (CZT) detector was evaluated. The detector had a field of view of 20 x 20 cm with detector elements of 2.5 x 2.5 mm in size. Both short-bore (35 mm) and long-bore (50 mm) collimators, matched to the geometry of the detector elements, were evaluated. The imaging performance of the CZT detector was compared with that of a conventional gamma camera equipped with all-purpose and ultra-high-resolution collimators. The performance of both systems with respect to breast imaging was evaluated using a water tank containing small glass spheres, 1.8-9.8 mm in diameter. The effects of variations in breast thickness, tumor depth, and tumor-to-background ratio were all simulated in this phantom model. Total counts per image were adjusted to approximate the count density observed in clinical scintimammographic studies.
Results:
Sensitivity of the CZT detector was 76% that of the equivalent NaI system. The system demonstrated excellent integral uniformity. The energy resolution of the CZT system was 6.5% for (99m)Tc. Spatial resolution with the long-bore collimator was superior to that of a conventional large field-of -view gamma camera equipped with an ultra-high- resolution collimator, over the range 0-6 cm from the collimator face. A blinded review of breast phantom images showed that small spheres (< or =7 mm in diameter) were better seen and had a better tumor-to-background ratio with the CZT system than with the conventional gamma camera.
Conclusion:
A small CZT detector offers superior performance to a conventional gamma camera and should permit reliable detection of breast tumors <1 cm in size.
More Related Videos
12:21Close-Space Sublimation-Deposited Ultra-Thin CdSeTe/CdTe Solar Cells for Enhanced Short-Circuit Current Density and Photoluminescence
Published on: March 6, 2020
11:26Integrating a Triplet-triplet Annihilation Up-conversion System to Enhance Dye-sensitized Solar Cell Response to Sub-bandgap Light
Published on: September 12, 2014
Related Concept Videos
Determination of Crystal Structures
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT