Related Experiment Video
Updated: Sep 26, 2026

Continuous Telemetric In Utero Tracheal Pressure Measurements in Fetal Lambs
Published on: December 22, 2023
[Nasal continuous positive airway pressure in preterm infants: comparison of two low-resistance models]
A Gutiérrez Laso1, P Sáenz González, I Izquierdo Macián
1Unidad de Cuidados Intensivos. Servicio de Neonatología. Hospital Universitario La Fe. Valencia. España. aguti@teleline.es
Objective:
To compare the safety and efficacy of two low expiratory resistance models of nasal continuous positive airway pressure (n-CPAP) in preterm infants.
Material And Methods:
A 1-year prospective trial was performed in the Neonatal Intensive Care Unit of La Fe Hospital to compare the Infant Flow (IF) and Medijet (MJ) devices. All preterm infants requiring n-CPAP for respiratory distress at birth (group I), infants weighting less than 1500 g requiring postextubation (group II) and those with apnea-bradycardia syndrome (ABS) (group III) were included. The patients were randomly assigned to IF or MJ.
Results:
A total of 125 patients received 226 treatments (IF: n 5 126: MJ: n 5 110). The mean gestational age was 29.4 weeks and the mean birth weight was 1340 g.Efficacy. In group I (n 5 73) no difference were found between systems and 6 hours' after initiation of n-CPAP decreases in FiO2, CO2 and respiratory effort were similar. The need for intubation was also similar (IF: 34.6 %; MJ: 24.1 %). In group II (n 5 73) the need for reintubation at 48 hours was similar with both treatments (IF:19 %; MJ: 8 %). In group III (n 5 80) resolution of ABS was similar after 24 hours of n-CPAP (IF: 46 %; MJ: 58 %). The need for intubation was also similar (IF: 26 %; MJ: 10 %).Complications. Air leaks occurred in six preterm infants (IF: 4; MJ: 2). Severe abdominal distension occurred in 5 % with both systems. Five infants had significant nasal lesions (IF: 1; MJ: 4).
Conclusions:
The efficacy and safety of both systems was similar in the variables studied and no significant differences were found.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation II: ACLS Airway Management
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...