A metal dispersed sol-gel biocomposite amperometric glucose biosensor
D Ravi Shankaran1, N Ueheara, T Kato
1Satellite Venture Business Laboratory, Utsunomiya University, Yoto 7-1-2, 321-8585, Utsunomiya, Japan.
Abstract:
A new glucose biosensor has been fabricated by immobilizing glucose oxidase into a copper dispersed sol-gel derived ceramic-graphite composite. The copper in the biocomposite offers excellent electrocatalytic activity towards the reduction (at -0.2 V) as well as oxidation (at +0.45 V) of hydrogen peroxide liberated in the enzymatic reaction enabling sensitive and selective determination of glucose. A linear response to glucose in the concentration range between 2.7 x 10(-5) to 4.0 x 10(-3) M with a correlation coefficient of 0.9987 and 4.0 x 10(-5) to 5.6 x 10(-3) M with a correlation coefficient of 0.9989 were observed with the electrocatalytic reduction and oxidation, respectively. Ascorbic acid and uric acid did not interfere with the glucose measurement during catalytic reduction at -0.2 V, a Nafion membrane was used to eliminate these interferences during the electrocatalytic oxidation at +0.45 V. The combination of copper catalysis and the promising feature of sol-gel biocomposite favor the sensitive and selective determination of glucose with improved analytical capabilities.