Multiple-use atomizers in outpatient otolaryngology clinics are not necessarily an infectious risk

Ann Marie B Visosky1, Andrew H Murr, Valerie Ng

  • 1University of California San Francisco School of Medicine, University of California San Francisco, USA.

Abstract

Insights

Multiple-use atomizers in otolaryngology clinics are safe for patient use. Bacterial contamination of phenylephrine and tetracaine solutions was minimal, posing no significant infectious risk.

Area of Science:

  • Otolaryngology
  • Infectious Disease Control
  • Microbiology

Background:

  • Multiple-use atomizers are common in outpatient otolaryngology settings.
  • Concerns exist regarding potential bacterial transmission via shared medical devices.

Purpose of the Study:

  • To assess the risk of bacterial contamination and transmission from multiple-use atomizers in an otolaryngology clinic.

Main Methods:

  • Bacterial cultures were performed on atomizers, associated tubing, and multidose phenylephrine/tetracaine solutions over two weeks.
  • Serial dilutions were used to overcome potential inhibitory effects of drug preservatives.

Main Results:

  • No bacterial growth was detected in cultures of the atomizer tubing.
  • Only 0.6% of drug solution cultures (2 out of 336) showed bacterial growth (coagulase-negative Staphylococcus).

Conclusions:

  • Phenylephrine and tetracaine solutions in multiple-use atomizers contain bacteriostatic preservatives.
  • These atomizers, when used appropriately, do not present a significant risk for bacterial transmission between patients.