Failure on Everest: the oxygen equipment of the spring 1952 Swiss expedition
1Department of Medicine, University of California San Diego, La Jolla, CA 92093-0623, USA. jwest@ucsd.edu
Abstract:
The Swiss expedition to Everest (altitude 8,848 m) in the spring of 1952 very nearly attained the summit with two of the climbers reaching 8,595 m. This would have been the first ascent of Everest. One main reason for failure was the inadequate oxygen equipment which was modified from sets designed for mine rescue using potassium peroxide generators. Oxygen could not be inhaled while the mountaineers were climbing because of the rigid mouthpiece assembly and the high breathing resistance of the equipment. It was argued that breathing oxygen during rest periods would suffice, but this was a fallacy as had been demonstrated by Kellas 30 years before.
More Related Videos
Related Concept Videos
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Pneumothorax-I
Pneumothorax can be even further classified as spontaneous, traumatic, and tension pneumothorax.
Pneumothorax-II
Clinical Manifestations:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...