Related Experiment Video
Updated: Sep 26, 2026

Simple Continuous Glucose Monitoring in Freely Moving Mice
Published on: February 24, 2023
A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 1)
A Poscia1, M Mascini, D Moscone
1Research and Development Department, A. Menarini Industrie Farmaceutiche Riunite S.r.l., Via Sette Santi, Florence, 3-50131, Italy. aposcia@menarini.it
Abstract:
The performances and the stability of a novel subcutaneous glucose monitoring system have been evaluated. GlucoDay (A. Menarini I.F.R. S.r.l, Florence Italy) is a portable instrument provided with a micro-pump and a biosensor coupled to a microdialysis system capable of recording the subcutaneous glucose level every 3 min. Long and short term stability of the biosensor are discussed and the results of some critical in vitro and in vivo (on rabbits) experiments are reported. A linear response up to 30 mM has been found for in vivo glucose concentration. The sensitivity referred to blood glucose is better than 0.1 mM and the zero current is typically below the equivalent of 0.1 mM. In the accuracy study a mean bias of 2.7 mg/dl and a correlation coefficient equal to 0.9697 have been found. At room temperature, an excellent membrane stability assures good performances up to 6 months from the first use.