Liquid chromatographic separation and determination of aromatic sulfonates in an aquatic environment using a
R Nageswara Rao1, N Venkateswarlu, Sara Khalid
1HPLC Group, Division of Analytical Chemistry, Discovery Laboratory, Indian Institute of Chemical Technology, Tarnaka, Hyderabad-500 007, India.
Abstract:
A simple and rapid method involving high-performance liquid chromatographic separation, followed by photodiode array (PDA) and electrospray ionization-mass spectrometric (ESI-MS) detection of aromatic sulfonates in waste-water effluents of industrial units producing optical whitening agents, has been developed. The separation was achieved on a reversed-phase Hypersil C18 column using gradient elution of a mobile phase consisting of 0.05 M ammonium formate-methanol with decreasing concentration of the buffer at room temperature. The minimum detection limits were determined to be in the range of 0.2 - 1.8 x 10(-9) g using PDA and ESI-MS detectors.
More Related Videos
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
09:36Applications of Liquid-Chromatography Tandem Mass Spectrometry in Natural Products Research: Tropane Alkaloids as a Case Study
Published on: March 8, 2024
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometry: Aromatic Compound Fragmentation
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...