Related Experiment Video
Updated: Aug 10, 2026

Iridium(III) Luminescent Probe for Detection of the Malarial Protein Biomarker Histidine Rich Protein-II
Published on: July 7, 2015
A novel 1,10-phenanthroline-sensitive membrane sensor for potentiometric determination of Hg(II) and Cu(II) cations
Sayed A M Marzouk1, Wadie T Al-Ariqui, Saad S M Hassan
1Department of Chemistry, Faculty of Science, Ain Shams University, Cairo, Egypt.
Abstract:
Preparation, characterization, and applications of a 1,10-phenanthrolinium cation (phenH(+))-sensitive potentiometric sensor are described. The sensor incorporates a liquid polymeric membrane consisting of phenH-tetraphenylborate, nitrophenyloctyl ether, and poly(vinyl chloride) as ion exchanger, plasticizer, and polymeric support, respectively. The sensor exhibits a fast and Nernstian response to phenH(+) over the concentration range of 6 x 10(-6)-2 x 10(-4) M with a monovalent cationic slope of 58.0+/-0.5 mV/log[phenH(+)] in acetate buffer of pH 4.2. The sensor is successfully applied to the monitoring of the potentiometric titration of Hg(II) and Cu(II) ions with phen solution in the presence of citrate and acetate buffers of pH 4.2, respectively. Sharp inflection breaks (90-180 mV) at 1:1 (metal:phen reaction) are obtained in the presence of chloride and thiocyanate background. This stoichiometry is explained by the formation of insoluble [HgCl(2)(phen)], [Hg(SCN)(2)(phen)], and [Cu(SCN)(2)(phen)] complexes. Optimization of each titration and the effect of foreign ions are evaluated. The method offers the advantages of adequate sensitivity, accuracy, and selectivity for the determination of mercury and copper in pharmaceutical, rock, and tea samples. The results are in good agreement with those obtained using the standard atomic absorption spectrometric and United States Pharmacopeial methods.
Related Concept Videos
Potentiometry: Overview
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Extraction: Advanced Methods