[Follow up of total hip prothesis: storage phosphor versus screen-film radiography]
K Polet1, H de La Selle, J Leroux
1Service d'imagerie Guilloz, Hôpital Central, 54035 Nancy-cedex.
Purpose:
Digital radiography may lead to interpretation difficulties in patients with hip prosthesis, especially when initial evaluation was performed using screen film radiography. The purpose of this study was to evaluate inter-observer and inter-technique reproducibility comparing digital and screen-film radiography.
Material And Methods:
Twenty-eight patients with unilateral (n=24) or bilateral (n=4) hip prosthesis (total=32 prostheses) underwent screen-film and digital (storage phosphor) radiography. The bone-cement and metal-cement interfaces were evaluated for the presence of a lucent line in 14 different sectors reclassified into 3 regions. This evaluation was performed by two independent readers. Inter-observer and inter-technique reproducibility were calculated using kappa values.
Results:
The inter-observer reproducibility was negligible, poor or average with Kappa values ranging from 0.07 to 0.47. The degree of agreement between the two methods was poor, average or good for both observers with Kappa values ranging from 0.10 to 0.66.
Conclusion:
Although inter-technique agreement is not high, it is better than inter-observer agreement. Therefore, storage phosphor technique should not be rejected for total hip prosthesis follow-up, even though previous films were screen-film radiographs.