[Low-dose chest CT with millimetric thin slices: myth or reality?]
O Corneloup1, O Delval, F Laurent
1Service de radiologie, CHU de Poitiers, Hôpital Jean Bernard, 2 rue de la Milétrie, 86000 Poitiers. olivier.corneloup@free.fr
Purpose:
To evaluate the diagnostic accuracy of millimetric thin slices low dose chest CT.
Materials And Methods:
Forty one patients underwent a chest CT thin slices (1 mm every 10 mm) exploration using both a 170 milliamperage and a low dose acquisition using 80 mA. The examination were read by 2 senior radiologists specialized in chest imaging without knowledge of acquisition parameters and in a random order. A statistical analysis of interobserver agreement was performed using Kappa analysis. Doses of both acquisition were estimated by compagning the dose length product calculated by the CT software and be using a simulation software.
Results:
Excellent interobserver and intermodalities agreements were found. A 53% decrease in dose was estimated with the low dose modality compare to the normal dose.
Conclusion:
Low dose thin slice chest CT using 80 mA has a similar diagnosis accuracy as standard dose thin slice chest CT and delivers half dose of irradiation.