Related Experiment Videos
Photo-elastic stress analysis of initial alignment archwires
Serene A Badran1, John F Orr, Mike Stevenson
1Orthodontic Division, School of Clinical Dentistry, Queen's University Belfast, Northern Ireland.
Abstract:
Photo-elastic models replicating a lower arch with a moderate degree of lower incisor crowding and a palatally displaced maxillary canine were used to evaluate the stresses transmitted to the roots of the teeth by initial alignment archwires. Six initial alignment archwires were compared, two multi-strand stainless steel wires, two non-super-elastic (stabilized martensitic form) nickel titanium wires, and two stress-induced super-elastic (austenitic active) nickel titanium wires. Three specimens of each archwire type were tested. Analysis of the photo-elastic fringe patterns, in the medium supporting the teeth, revealed that the non-super-elastic nickel titanium archwires produced the highest shear stresses (P = 0.001). However, the shear stresses generated by the super-elastic alignment archwires and the multi-strand stainless steel archwires were very similar (P = 1.00). These results show that even in situations where large deflections of initial alignment archwires are required, super-elastic archwires do not appear to have any marked advantage over multi-strand stainless steel alignment archwires in terms of the stresses transferred to the roots of the teeth.