Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography
1Department of Cardiology, Klinikum Grosshadern, University of Munich, Munich, Germany. alexander.leber@med1.med.uni-muenchen.de
Background:
Electron beam computed tomography (EBCT) and multislice computed tomography (MSCT) are both suitable for non-invasive identification of coronary stenoses.
Objective:
To compare intravenous coronary EBCT angiography (EBCTA) and MSCT angiography (MSCTA) with regard to image quality and diagnostic accuracy.
Methods:
EBCTA was done using an Imatron C-150 XP scanner in 101 patients following a standard protocol (slice thickness 3 mm, overlap 1 mm, acquisition time 100 ms, prospective ECG trigger). For MSCTA in a different set of 91 patients (using a Siemens Somatom Plus4VZ scanner), the whole volume of the heart was covered in a spiral technique by four simultaneous detector rows. Using retrospective ECG gating, the raw data were reconstructed in (mean (SD)) 215 (12) axial slices acquired in diastole (slice thickness 1.25 mm, overlap 0.5 mm, acquisition time 250 ms/slice).
Results:
With EBCTA, 76% of predetermined coronary segments in a nine segment model could be assessed with diagnostic image quality, and with MSCTA, 82%. A low contrast to noise ratio with EBCTA, and the presence of motion artefacts with MSCTA were the main reasons for inadequate image quality. Using conventional angiography as the gold standard, 77% of stenoses of > 50% could be identified correctly with EBCTA and 82% with MSCTA. Significant stenoses were correctly ruled out in 93% of segments with EBCTA, and in 96% of segments with MSCTA. The average contrast to noise ratio was higher with MSCTA than with EBCTA (9.4 v 6.5; p < 0.001).
Conclusions:
EBCTA and MSCTA show similarly high levels of accuracy for determining and ruling out significant coronary artery stenoses. MSCTA is capable of providing good image quality in more coronary segments than EBCTA because of its better contrast to noise ratio and higher spatial resolution. Motion artefacts seen at heart rates of > 75 beats/min and a higher radiation exposure are the main limitations of MSCTA.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies III: Computed Tomography