Development of an In Vitro Ocular Platform to Test Contact Lenses
Published on: April 6, 2016
Oxygen permeability of the hypertransmissible contact lenses
Matthew D Young1, William J Benjamin
1School of Optometry, University of Alabama at Birmingham Birmingham, AL 35294-0010, USA.
Purpose:
Oxygen permeability (Dk) was assayed in a masked fashion using the polarographic method for three materials found capable of hypertransmissibility as contact lenses: tisilfocon A, balafilcon A, and lotrafilcon A.
Methods:
Two sets of tisilfocon rigid lenses were tested in six thicknesses and two sets in nine thicknesses for the hydrogels. The two sets of tisilfocon were from two different lots and of lotrafilcon were from different base curve radii (8.6 and 8.4 mm). The six test sets were measured with the inclusion of rigid reference materials for the purpose of simultaneous calibration.
Results:
Oxygen permeability (in barrers) of the two lots of tisilfocon A was 159.6 (95% confidence interval [CI], 146.3-175.5) and 175.1 (95% CI, 154.9-201.6). The Dk values for balafilcon A were 111.3 (95% CI, 103.6-120.3) and 102.0 (95% CI, 90.5-116.7), and the Dk values for lotrafilcon A were 176.1 (95% CI, 165.1-188.7) and 190.2 (95% CI, 177.5-205.0) The coefficients of determination (R2) of the corrected resistance versus thickness plots were 0.98 or more. Paired-comparison t tests were performed between the sets of each test material paired by thickness. These revealed the oxygen transmissibilities (Dk/L) of the lenses made from the two lots of tisilfocon A to be significantly different (t = 2.62; P < 0.047; df = 5) and lenses of the two base curve radii made of lotrafilcon A to be significantly different (t = 4.42; P < 0.002; df = 8). These differences were attributed to a material difference in Dk between lots of tisilfocon A and a thickness difference between lens designs of lotrafilcon A.
Conclusion:
Oxygen permeability of tisilfocon was found consistent with that claimed by the manufacturer (163 barrers). Dk of balafilcon was found marginally greater than claimed (99 barrers) and of lotrafilcon, greater than claimed (140 barrers). Preliminary Dk/L values of two of the three test materials were statistically different between Sets 1 and 2, yet they were inside the American National Standards Institute Z80.20-1998 tolerance (+/-20%). Resistance-versus-thickness plots were linear up to a Dk of at least 190 barrers. The Dk/L values over the central zone of minus and low-plus contact lenses made of these materials are predicted to decrease on the upper asymptotic portion of the Dk/L versus equivalent oxygen percentage curve. Hence, the physiologic effects of oxygen variability from lot-to-lot, lens-to-lens, and site-to-site on the same lens will be minimized.
Related Concept Videos
Oxygen Transport in the Blood
Oxygen Requirements and Growth Patterns
External and Internal Respiration
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ophthalmic Drug Delivery Systems