A reagentless nitric oxide biosensor based on haemoglobin/polyethyleneimine film
Related Experiment Video
Updated: Sep 25, 2026

Analytical Techniques for Assaying Nitric Oxide Bioactivity
Published on: June 18, 2012
Xinjian Liu1, Libin Shang, Jiantao Pang
1Department of Biochemistry and National Key Laboratory of Pharmaceutical Biotechnology, Nanjing University, Nanjing 210093, People's Republic of China.
Abstract:
An unmediated NO biosensor was prepared by co-immobilizing haemoglobin with polyethyleneimine at a pyrolytic graphite electrode surface. Haemoglobin exhibits direct electrochemistry in the membrane, and this modified electrode shows high stability and catalytic activity towards the reduction of NO. A linear relationship between the catalytic current and the concentration of NO was observed in the concentration ranges 1.0 x 10(-7)-8.0 x 10(-6) M and 1.0 x 10(-5)-2.5 x 10(-5) M, with a detection limit of 4.0 x 10(-8) M.