Inflation of a pressure-limited cuff inside a model trachea
Related Experiment Video
Updated: Sep 21, 2026

Ascending Aortic Constriction in Rats for Creation of Pressure Overload Cardiac Hypertrophy Model
Published on: June 29, 2014
1Intensive Care Unit, Queen Elizabeth Hospital, King's Lynn, Norfolk PE30 4ET, UK.
Abstract:
A theory of the elastic behaviour of thin rubber films is developed to describe inflation of pressure-limited tracheal tube cuffs (a) outside and (b) inside rigid cylindrical model tracheas. At each stage, assumptions and results were successfully checked against experiment. The theory predicts cylinder wall versus intracuff pressure. This allows, for any suitable cuff (of which there are many), the determination of a single intracuff pressure that ensures a satisfactory seal, for any diameter within the human adult range, the wall pressure staying below some chosen safety limit.
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...