Measurement of cerebral oxygenation state in anaesthetized children using the INVOS 5100 cerebral oximeter
Alexander Dullenkopf1, Bernhard Frey, Oskar Baenziger
1Departments of Anaesthesia and Intensive Care and Neonatology, University Children's Hospital Zurich, Steinwiesstrasse 75, CH-8032 Zurich. markus.weiss@kispi.unizh.ch
Background:
Near-infrared spectroscopy is a developing technique for monitoring cerebral oxygenation during anaesthesia. The aim of this study was to evaluate absolute values of cerebral oxygenation during stable anaesthesia conditions in otherwise healthy children using the recently introduced INVOS 5100 cerebral oximeter with a paediatric and adult sensor and to compare them with values obtained from the NIRO 300 oximeter.
Methods:
Thirty paediatric surgical patients (aged 0.23-15.97 years) were studied during general anaesthesia with tracheal intubation and controlled ventilation. Comparative measurements of cerebral oxygenation were performed on the forehead with two probes within 10 min under stable cardiorespiratory and anaesthesia conditions. Cerebral oxygenation values (rSO2) obtained from the paediatric and adult INVOS 5100 sensors were compared with the tissue oxygenation index (TOI) obtained from the NIRO 300 cerebral oximeter using 4- and 5-cm emitter-detector separation.
Results:
Cerebral rSO2 values and the TOI values both showed a large range of cerebral oxygenation in the children studied (rSO2: 59-95%, TOI: 48-85%). Cerebral rSO2 values measured by the INVOS 5100, particularly with the paediatric sensor, were significantly higher than the TOI values obtained from the NIRO 300 (P < 0.0001). Agreement between the INVOS and NIRO oximeter was poor.
Conclusion:
The large range and the poor agreement of cerebral oxygenation values between the two oximeters makes it difficult to define a normal value. Cerebral oxygenation readings by these monitors, based on one single point measurement during anaesthesia, should be viewed with caution. Actually, there may be little indication for routine use of such monitoring during general anaesthesia.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...