Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models
Published on: November 11, 2020
Albuterol delivery with conventional and synchronous ventilation in a neonatal lung model
Sandra S Garner1, W Michael Southgate, Donald B Wiest
1Department of Pharmaceutical Sciences, Medical University of South Carolina, Charleston, SC, USA.
Objective:
To compare the percentage of nebulized albuterol delivered with conventional (intermittent mandatory ventilation) vs. synchronous (assist-control and assist-control with flow synchronization) ventilation in a neonatal lung model.
Design:
Prospective in vitro laboratory study.
Setting:
Research laboratory.
Subject:
Neonatal lung model.
Interventions:
The model simulated an intubated neonate with a spontaneous respiratory rate of 40, 60, or 80 breaths per minute and compliance and resistance values of bronchopulmonary dysplasia. A VIP Bird ventilator was used for all ventilator modes. Albuterol 2.5 mg was administered with a T Up-Draft II Neb-U-Mist nebulizer attached to a 12.75-cm (10-mL) reservoir of circuit tubing. Albuterol was collected onto a filter (particle retention
Measurements And Main Results:
No significant differences in percentage albuterol delivered were found among the three modes or the three spontaneous respiratory rates (mean +/- sd): intermittent mandatory ventilation, 0.11 +/- 0.04%; assist-control, 0.12 +/- 0.03%; assist-control with flow synchronization, 0.10 +/- 0.04%; 40 breaths per minute, 0.11 +/- 0.03%; 60 breaths per minute, 0.11 +/- 0.04%; and 80 breaths per minute, 0.11 +/- 0.05% (p >.05, two-factor analysis of variance).
Conclusions:
The percentage of nebulized albuterol delivered to the end of the endotracheal tube in a mechanically ventilated neonatal model was not affected by mode of ventilation under the conditions studied. Further clinical studies are needed to determine whether lung deposition and distribution or clinical efficacy of albuterol is influenced by synchronous ventilation in this patient population.
More Related Videos
Related Concept Videos
Pulmonary Cycle: Exhalation
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management