Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?
Christof Kolb1, Sebastian Schmieder, Günter Lehmann
1Deutsches Herzzentrum München, Klinikum rechts der Isar, Elektrophysiologie, Technischen Universität München, Lazarettstrasse 36, 80636 Munich, Germany. Kolb@dhm.mhn.de
Objectives:
The aim of this study was to determine whether airport metal detector gates (AMDGs) interfere with pacemakers (PMs) or implantable cardioverter-defibrillators (ICDs).
Background:
It is currently unknown whether AMDGs interfere with implanted PMs or ICDs.
Methods:
A total of 348 consecutive patients (200 PM and 148 ICD recipients) have been tested for the occurrence of electromagnetic interference (EMI) within the electromagnetic field of a worldwide-used airport metal detector.
Results:
No interference, such as pacing or sensing abnormalities, was observed in any of the 200 PM and 148 ICD patients; also no reprogramming occurred.
Conclusions:
In vivo testing of PM and ICD systems showed no EMI with a standard AMDG. Clinically relevant interactions with implanted PMs or ICDs seem unlikely.
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Imaging Studies for Cardiovascular System IV: CMRI
Holter Monitor: 24-Hour Monitoring
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias