Use of an automated blood pressure recording device, the BpTRU, to reduce the "white coat effect" in routine practice
Related Experiment Video
Updated: Aug 21, 2026

Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
Martin G Myers1, Miguel A Valdivieso
1Division of Cardiology, Schulich Heart Centre, Sunnybrook and Women's College Health Sciences Centre, Toronto, Ontario, Canada.
Background:
Patients often exhibit higher blood pressure (BP) readings in the doctor's office, a phenomenon known as the white coat effect. This study examines the presence of a physician in the examining room as a possible factor in provoking a white coat effect.
Methods:
Blood pressure measurements taken by an automated BP recording device, the BpTRU (VSM MedTech Ltd., Vancouver, BC, Canada) with the patient alone in the examining room, were compared with the following: (1) BP taken by the patient's family physician; (2) BP taken on the first visit to a hypertension specialist; (3) BP measured by a trained research technician and (4) the mean awake ambulatory BP (ABP). The BpTRU and trained research technician readings were taken outside of the office (treatment) setting in an ABP research unit.
Results:
Blood pressure readings (mm Hg, mean +/- SEM) taken by the BpTRU (155 +/- 5/88 +/- 2) tended to be lower than for the family physician (166 +/- 4/89 +/- 3) and the hypertension specialist (174 +/- 5/92 +/- 2; P <.001). However, BP taken by the trained research technician (158 +/- 4/90 +/- 2) was similar to the value obtained by the BpTRU. The mean awake ABP was lower (P < 0.01) than the other four BP values.
Conclusions:
Use of an automated BP recording device outside of the office (treatment) setting can partly eliminate the white coat effect. A similar finding was observed with readings taken by a trained research technician under similar conditions. Referral of patients to nonoffice settings for automated BP recordings may provide a more accurate estimate of a patient's BP status, with partial elimination of the white coat effect associated with readings taken by a physician.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure