Visual verification of linac light and radiation fields coincidence
Angelo F Monti1, Milena Frigerio, Giovanna Frigerio
1Department of Medical Physics, Como, Italy. amonf@iol.it
Abstract:
X-ray and light field alignment evaluation is carried out during linac quality assurance programs. In this paper, we compare the size of the light field measured by a photodiode and by a more traditional visual observation with the size of the x-ray field. The comparison between actual light field size, measured with the photodiode, and light field size measured by human eye allow us to verify the reliability of human eye in the evaluation of this parameter. The visual field is always larger than real light field; however, it agrees better with the x-ray field. It matches the light field if we take into account the 25% (+/- 1%) of the decrement line of the maximum central lightening; however, this method simulates better the actual field employed in radiation treatments.