Nuchal translucency thickness measurement: repeatability using a virtual ultrasound scanner
V R Newey1, D K Nassiri, A Bhide
1Department of Medical Physics and Bioengineering, St. George's Hospital, London, UK.
Summary
A PC-based virtual ultrasound scanner (VirUS) assessed nuchal translucency (NT) measurement repeatability. Results show VirUS offers a consistent environment, highlighting the need for optimized gain settings and consistent thickness measurements for improved accuracy.
Area of Science:
- Medical Imaging
- Fetal Ultrasound Technology
Background:
- Nuchal translucency (NT) measurement is crucial for fetal aneuploidy screening.
- Operator variability can impact the accuracy of NT measurements.
- A consistent and reliable method for assessing measurement repeatability is needed.
Purpose of the Study:
- To evaluate the inter- and intraoperator repeatability of nuchal translucency (NT) thickness measurements using a PC-based virtual ultrasound scanner (VirUS).
- To assess the performance of experienced ultrasound operators in a simulated NT measurement environment.
Main Methods:
- Simulated fetal ultrasound images with defined NT thickness were generated using VirUS.
- Thirteen experienced operators performed NT thickness measurements on a set of 50 simulated images across three separate occasions.
- A total of 1950 measurements were collected to assess repeatability.
Main Results:
- Interoperator repeatability was +/- 0.41 mm and intraoperator repeatability was +/- 0.22 mm (95% confidence level).
- Significant correlations were found between repeatability and mean measured NT thickness, as well as between scanner gain settings and measured thickness.
- Operator performance varied, with some showing significant correlations between gain and measurement outcomes.
Conclusions:
- The virtual ultrasound scanner (VirUS) provides a consistent environment for auditing NT measurements.
- Optimizing repeatability in relation to mean measured thickness and ensuring consistent gain settings are critical for accurate NT assessment.
- VirUS demonstrates potential as a tool for operator training and quality assurance in NT measurement.