Related Experiment Video
Updated: Sep 25, 2026

Automated Hospital Room Disinfection Utilizing a Novel Aerosolized Hydrogen Peroxide Microdroplet Disbursing Technology
Published on: February 24, 2026
Endoscope disinfection by ortho-phthalaldehyde in a clinical setting: an evaluation of reprocessing time and costs
1Endoscopy Unit, Massachusetts General Hospital, 55 Fruit Street, Boston, MA 02114, USA. shession@partners.org
Abstract:
Ortho-phthalaldehyde provides high-level disinfection in 12 minutes at room temperature (20 degrees C). To determine whether the endoscopy unit at Massachusetts General Hospital could cost-effectively increase endoscope throughput and decrease delayed procedures while maintaining biocidal efficacy with heavy use, ortho-phthalaldehyde solution (0.55% w/v) was used for high-level disinfection in place of glutaraldehyde in the unit's five automatic endoscope reprocessors. During this 38-day study, 1,839 endoscopes were processed in 1,201 cycles. Each machine ran an average of 80 disinfection cycles in each of three consecutive periods. The shorter disinfection time of ortho-phthalaldehyde (12 minutes) resulted in an average savings of 53 hours in disinfection time per study period for the five processors when compared with glutaraldehyde's 20-minute disinfection time.Ortho-phthalaldehyde costs more per gallon than glutaraldehyde ($31.03 versus $13.20). This higher cost would be offset, however, by additional time savings resulting from the fact that compared with glutaraldehyde, ortho-phthalaldehyde is faster-acting and a mixing and activation step is not required. In addition, testing prior to each cycle verified that despite heavy use, ortho-phthalaldehyde solution remained efficacious, lasting through an average of 80 cycles, whereas glutaraldehyde only lasts for an average of 40 cycles. Test strips showed the concentration of ortho-phthalaldehyde in the reprocessors remained above the minimum effective concentration for the entire 14-day maximal reuse period. Additional microbiological efficacy testing of spent solution diluted to its minimum effective concentration demonstrated the solution remained tuberculocidal in tests with Mycobacterium bovis.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...