Body contour 180 degrees pinhole SPET with or without tilted detector: a phantom study
Alain Seret1, David Flérès, Olivier Firket
1Imagerie médicale expérimentale et Centre de Recherche du Cyclotron, Institut de Physique, Université de Liège, Liège 1, Belgium. aseret@ulg.ac.be
Abstract:
This study investigated the feasibility of ordered subsets expectation maximisation (OS-EM) reconstruction of 180 degrees pinhole single-photon emission tomography (SPET) acquired in body contour mode (variable distance between the detector and the axis of rotation for each projection) with or without a tilted detector head. Four non-circular orbits were designed bearing in mind the rotation radius and tilt angle values of previous pinhole SPET acquisitions in patients with circular orbits. The reconstructions were performed using a dedicated OS-EM algorithm. Reconstructed images of line and uniformity phantoms showed that the spatial and uniformity characteristics of the radioactive objects were preserved. In comparison with the circular orbits, the non-circular orbits allowed only a moderate gain (maximum 10%) in resolution. However, body contour pinhole SPET would significantly facilitate the camera set-up and in this way should decrease the camera set-up time, which is an important parameter in patient studies.