[Dose reduction through gridless technique in digital full-field mammography].

F Diekmann1, S Diekmann, S Berzeg

  • 1Institut für Radiologie, Universitätsklinikum Charité, Berlin. felix.diekmann@charite.de

Rofo : Fortschritte Auf Dem Gebiete Der Rontgenstrahlen Und Der Nuklearmedizin
|June 18, 2003
PubMed
Summary

Gridless digital mammography offers improved image quality and dose reduction for thin breasts. Further adjustments are needed for overall dose reduction in gridless systems.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Dynamic contrast enhanced MRI of the head and neck region using a VIBE sequence with Cartesian undersampling and compressed sensing.

Magnetic resonance imaging·2024
Same author

Momentum-space signatures of Berry flux monopoles in the Weyl semimetal TaAs.

Nature communications·2021
Same author

Spectral CT in clinical routine imaging of neuroendocrine neoplasms.

Clinical radiology·2021
Same author

Switching off for future-Cost estimate and a simple approach to improving the ecological footprint of radiological departments.

European journal of radiology open·2021
Same author

Inter- and Intrareader Agreement of NI-RADS in the Interpretation of Surveillance Contrast-Enhanced CT after Treatment of Oral Cavity and Oropharyngeal Squamous Cell Carcinoma.

AJNR. American journal of neuroradiology·2020
Same author

Prostate Cancer Nodal Staging: Using Deep Learning to Predict <sup>68</sup>Ga-PSMA-Positivity from CT Imaging Alone.

Scientific reports·2020

Area of Science:

  • Radiology
  • Medical Imaging
  • Mammography

Background:

  • Scatter grids are used in mammography to improve image quality by reducing scattered radiation.
  • The role of scatter grids in digital full-field mammography regarding image quality and radiation dose requires further investigation.

Purpose of the Study:

  • To evaluate the impact of scatter grids on image quality and radiation dose in digital full-field mammography.
  • To compare experimental findings with initial clinical experiences.

Main Methods:

  • Utilized a CD-Mam phantom with varying gold dot sizes and thicknesses to simulate breast tissue.
  • Performed digital mammography with and without a scatter grid, simulating four breast thicknesses using Plexiglas.
  • Evaluated 16 mammograms by three readers, comparing image quality, radiation dose, and breast thickness.

Related Experiment Videos

Main Results:

  • Gridless digital mammography demonstrated superior image quality for simulated thin breasts at standard entrance doses.
  • Image quality differences between grid and gridless systems were negligible for simulated large breasts.
  • Gridless mammography showed greater advantages with a reduced entrance dose, achieving identical parenchymal doses.

Conclusions:

  • Gridless digital mammography presents an opportunity for further radiation dose reduction.
  • System adjustments, specifically a higher "switch-off dose" at the detector, are necessary for gridless systems to achieve overall dose reduction.
  • Gridless mammography may be inferior to grid mammography at identical detector plane doses, necessitating optimization.