Loaned self-measurement equipment model compared with ambulatory blood pressure monitoring
Calvo-Vargas César1, Padilla-Rios Victoria, Troyo-Sanromán Rogelio
1University of Guadalajara Department of Internal Medicine, Hospital Civil Juan I. Menchaca, Mexico. mrrm@prodigy.net.mx
Objective:
To compare the correspondence between the loaned self-measurement equipment model (LSEM), 24 h awake ambulatory blood pressure monitoring (ABPM) and clinic blood pressure (CBP).
Design:
A cross-sectional study was carried out involving 90 hypertensive and normotensive subjects.
Methods:
We used a variation of home measurement, the loaned self-measurement equipment model (LSEM), that consisted of lending a number of sphygmomanometers, the property of the clinic, to patients for 3-day periods. The correspondence of results from LSEM was measured against the difference between the two averages, the correlation coefficients and the average unweighted absolute differences.
Results:
The average difference between CBP and ABPM was 6.1/2.5 mmHg (P<0.001, r =0.54/0.63). Between CBP and LSEM, it was 6.6/1.4 mmHg (P<0.001/ 0.08, r =0.52/0.63), and between the ABPM and LSEM 0.5/-1.0 mmHg (P=0.70/0.14, r =0.71/0.75). The average unweighted absolute differences between LSEM and ABPM readings were 5.5 and 1.5 mmHg for systolic and diastolic blood pressure, respectively, whereas the average difference between CBP and ABPM means was much higher, at 6.8 mmHg for systolic and 2.7 mmHg for diastolic pressure. Agreement in the diagnosis of systolic hypertension between LSEM and ABPM was found in 78.8% of the individuals, and of diastolic hypertension in 76.7%.
Conclusion:
Blood pressure readings taken using LSEM have a better correspondence than CBP in predicting the awake ABPM value, although the disagreement in terms of diagnosing hypertension should be noted.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: