A system for integrated measurement of ventilator settings, lung volume change and blood gases during high-frequency
D G Markhorst1, H R van Genderingen, T Leenhoven
1Pediatric Intensive Care Unit, University Medical Center Utrecht, The Netherlands. dg.markhorst@vumc.nl
Abstract:
To describe and validate a system for integrated measurement of ventilator settings and dependent physiological variables during high-frequency oscillatory ventilation (HFOV). A custom interface was built for data acquisition. Lung volume change was determined by respirator inductive plethysmography (RIP), modified to sampling rates of 140 Hz. Blood gas analysis was obtained using a continuous intra-arterial blood gas monitoring system. FIO2 was measured by means of an electrochemical sensor. Pressure at the airway opening and trachea (microtip transducer) were sampled. The data acquired were sent to a laptop computer for analysis, display and storage. The system was tested during a lung recruitment procedure in an animal model of respiratory distress. Linearity of the RIP was checked by gas volume injection using a supersyringe. The system operated successfully. Agreement between RIP-measured volume with injected volume was excellent; bias was 5 ml; limits of agreement were 1-9 ml. Graphs were obtained, showing the relationship between imposed mean airway pressure and lung volume change, and oxygenation. The integration of ventilator settings and dependent physiological variables may provide useful information for clinical, instructional and research application.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)