Related Experiment Video
Updated: Sep 23, 2026

Fabrication of a Solution-gated Indium-Tin-Oxide-based One-piece Transistor Enabling Sensitive Biosensing
Published on: August 29, 2025
Monitoring trace anion contamination in disk drive components
Edward Kaiser1, Jeff Rohrer, Faye Campbell
1Dionex Corporation, PO Box 3603, 1228 Titan Way, Sunnyvale, CA 94088-3603, USA. edward.kaiser@dionex.com
Abstract:
Ion chromatography was used to determine trace anionic contamination on the surface of hard disk drive components. These contaminants can have a detrimental effect on device reliability and yield. Disk drive components were soaked in deionized water and these extracts were analyzed for anions. The anions fluoride, acetate, formate, acrylate, methacrylate, chloride, nitrite, bromide, nitrate, benzoate, sulfate, oxalate, phthalate and phosphate were separated on a high-performance anion-exchange column and determined at concentrations less than 1 microg/l with suppressed conductivity detection. The extract solutions were analyzed either by injecting 1 ml or by preconcentrating 5 ml. We evaluated the performance of both methods.