Related Experiment Video
Updated: Jul 29, 2026

A Simple Non-invasive Method for Temporary Knockdown of Upper Limb Proprioception
Published on: March 3, 2018
The KickStrip: a novel testing device for periodic limb movement disorder
Tamar Shochat1, Arie Oksenberg, Noam Hadas
1Scientific Laboratory Products, Ltd., Tel Aviv, Israel.
Study Objectives:
In light of the ongoing debate over the clinical significance of periodic limb movement disorder (PLMD) and its monitoring in overnight sleep studies, we introduce a novel, portable, low-cost device for PLMD testing. The KickStrip is a disposable device, which includes a movement sensor, a central processing unit with real-time software, and a display. In the present study, the KickStrip final score (Kscore) is validated against the traditional periodic limb movement index (PLMI) based on overnight recordings in the sleep laboratory. DEISGN: Patients underwent full polysomnographic recordings concomitantly with the use of the KickStrip for a single night.
Setting:
Sleep Disorders Unit at Loewenstein Rehabilitation Hospital and Sleep Medicine Center at Rambam Medical Center, Israel.
Patients:
Eighty-two patients referred for overnight sleep recordings due to sleep disturbance of any kind.
Interventions:
N/A.
Measurements And Results:
The Kscores were collected and the PLMI computed for each leg separately. Pearson correlations between Kscores and PLMI ranged between r = 0.83 to r = 0.88 (p < 0.001). Sensitivity and specificity values of the Kscores for increasing PLMD thresholds showed sensitivity ranging from 50% to 100% and specificity ranging from 83% to 100%. Receiver operating characteristic curves showed area-under-the-curve values ranging from 82% to 92%. Bland-Altman plot showed high agreement between the methods.
Conclusions:
Comparisons between the Kscores and the traditional PLMI show increased accuracy with severity level and excellent agreement. The KickStrip is a valuable tool for PLMD testing for both clinical and research purposes.