Inhaled nitric oxide (iNO) delivery with high-frequency jet ventilation (HFJV)
David R Platt1, Doug Swanton, David Blackney
1Bunnell Incorporated, Salt Lake City, UT 84115, USA.
Objective:
To determine if nitric oxide (NO) therapy can be reliably administered during high-frequency jet ventilation (HFJV) using the INOvent delivery system.
Study Design:
NO concentrations were measured just proximal to the endotracheal (ET) tube and at the distal tip of the ET tube during a bench evaluation. Measurements were taken over a wide range of airway pressure settings and NO concentrations with both high- and low- resistance lung models. Percent changes in set versus proximal and proximal versus distal iNO concentrations were tabulated.
Results:
Differences between proximal and distal NO concentrations were 10% or less. In the therapeutic range of up to 20 p.p.m., differences in concentration were 1 p.p.m. or less. There was no consistent effect on NO concentration when airway resistance was increased by 500%.
Conclusion:
Nitric oxide therapy can be reliably administered during HFJV with the INOvent delivery system when NO is injected exclusively via the HFJV circuit.