Use of Ultra-high Field MRI in Small Rodent Models of Polycystic Kidney Disease for In Vivo Phenotyping and Drug Monitoring
Published on: June 23, 2015
Inductively coupled rf coils for examinations of small animals and objects in standard whole-body MR scanners
Hansjörg Graf1, Petros Martirosian, Fritz Schick
1Section on Experimental Radiology, Department of Diagnostic Radiology, University of Tübingen, 72076 Tübingen, Germany. hansjoerg.graf@med.uni-tuebingen.de
Abstract:
Inductively coupled solenoid coils fitting to objects in the size of mice or rats were developed to adapt modem whole-body MR scanners featuring sufficient gradient strength for animal examinations with high spatial resolution. Homogenous receiver characteristics is achievable over almost the whole inner region of the solenoid coils. The SNR can be increased by a factor 2 to 6 with the adapting coils for examinations using the head coil as connected receiver. Standard sequences on clinical 1.5 T scanners can be applied with adapted transmitter voltages. For example, a SNR value of about 30 is achievable in a mouse liver after 10 minutes measuring time using a 2-D spin echo imaging sequence and a size of 0.3 x 0.3 x 0.8 mm3 for the picture elements.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI