Aim:
The purpose of this study was to compare the Cosmed K4b2 portable gas analysis system with the Cosmed Quark b2 metabolic cart.
Methods:
Twenty-one subjects attended one testing session that consisted of duplicate measurements of gas volumes and concentrations using both Cosmed gas analysis systems at 3 treadmill work rates; 1) 80m x min(-1), 0% grade, 2) 80m x min(-1), 5% grade, and 3) 80m x min(-1), 10% grade. Subjects walked for 3 min at each rate with one of the gas analysis systems attached to the facemask. The order of the procedures was randomized so that one system was used during both phases (1st or 2nd) of each work rate.
Results:
The results indicated that oxygen consumption (VO2) was significantly higher in the K4b2 compared to the Quark at 80m x min(-1), 0% grade (14.3+/-1.2 vs 13.6+/-1.2ml x kg(-1) x min(-1), respectively), (p<0.01). The fractional concentration of oxygen in expired air was also significantly lower in the K4b2 at 80 m x min(-1), 0% grade and 80 m x min(-1), 10% grade (p<0.05). There were no significant differences between systems for minute ventilation or carbon dioxide production. Despite the small mean bias in mean VO2 values (0.5-1.0ml x kg-1 x min(-1) higher) in the K4b2, all individual values were within the limits of agreement (mean difference+/-2 SD) as determined by the Bland-Altman technique.
Conclusion:
The findings show a minimal bias in respiratory and metabolic parameters during bi-pedal locomotor activities at low to moderate exercise intensities in the two gas analysis systems.
More Related Videos
05:02Comparing Bibliometric Analysis Using PubMed, Scopus, and Web of Science Databases
Published on: October 24, 2019
07:26Conducting Maximal and Submaximal Endurance Exercise Testing to Measure Physiological and Biological Responses to Acute Exercise in Humans
Published on: October 17, 2018
Related Concept Videos
Comparing the Survival Analysis of Two or More Groups
Comparative Excretory Systems
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Qualitative Analysis
For instance, group IV...
Gas Exchange and Transport
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...