Corneal thickness at different reference points from Orbscan II system.

Andrew K C Lam1, Judy S Chan

  • 1Department of Optometry and Radiography, The Hong Kong Polytechnic University, Kowloon, Hong Kong SAR, China.

Clinical & Experimental Optometry
|July 16, 2003
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Occult Retinal Abnormalities Detected by Routine Optical Coherence Tomography in Comprehensive Eye Examination: A Clinic Dataset.

Clinical optometry·2026
Same author

Changes of Schlemm's Canal and Trabecular Meshwork Dimensions from Digital Ocular Massage in High Myopia: A Pilot Study.

Ophthalmic research·2025
Same author

Short-term digital ocular massage may weaken corneal biomechanics.

Frontiers in bioengineering and biotechnology·2025
Same author

Real-world outcomes of Defocus Incorporated Multiple Segments lenses on retarding axial elongation in myopic children and adolescents.

Frontiers in medicine·2025
Same author

Performance of neuroretinal rim thickness measurement by Cirrus high-definition optical coherence tomography in myopic eyes.

Scientific reports·2023
Same author

Excessive use of electronic devices among children and adolescents is associated with musculoskeletal symptoms, visual symptoms, psychosocial health, and quality of life: a cross-sectional study.

Frontiers in public health·2023
Same journal

An international survey of low vision rehabilitation services.

Clinical & experimental optometry·2026
Same journal

Readiness of the tabletop screen-based perimetry tests for screening and monitoring glaucoma.

Clinical & experimental optometry·2026
Same journal

Validation of the vision quality of life-time survey: comparison with the convergence insufficiency symptom survey.

Clinical & experimental optometry·2026
Same journal

Measuring contrast sensitivity in school children and young adults: a comparison between Pelli-Robson and SpotChecks charts.

Clinical & experimental optometry·2026
Same journal

Three-year retinal and visual outcomes after ultraviolet- and blue light-filtering intraocular lenses.

Clinical & experimental optometry·2026
Same journal

Repeatability of choroidal optical coherence tomography angiography metrics in healthy young adults.

Clinical & experimental optometry·2026
See all related articles

The thinnest corneal region is typically inferior, but its location doesn't significantly differ in thickness from the corneal apex. Orbscan II users can confidently use central reference points for corneal thickness measurements.

Area of Science:

  • Ophthalmology
  • Corneal topography
  • Ocular biomechanics

Background:

  • Corneal topography systems like Orbscan II define reference points for analysis.
  • Understanding the precise location of the thinnest corneal region is crucial for accurate assessments.
  • Investigating potential discrepancies between the corneal apex and the thinnest corneal zone is important.

Purpose of the Study:

  • To determine the location of the thinnest corneal region using the Orbscan II system.
  • To assess if there is a significant difference in corneal thickness between the corneal apex and the thinnest corneal region.
  • To evaluate the clinical significance of reference point location variations.

Main Methods:

  • Analysis of 92 eyes from 46 healthy subjects using the Orbscan II system.

Related Experiment Videos

  • Identification and categorization of three reference points: fixation point, corneal apex, and thinnest corneal region.
  • Statistical analysis to compare the location and thickness at these reference points.
  • Main Results:

    • The thinnest corneal region was predominantly located in the inferior temporal quadrant (right eye) and inferiorly (left eye).
    • No statistically significant difference was observed in corneal thickness between the thinnest region and the corneal apex.
    • A significant difference (p < 0.01) was found in the spatial location of these identified reference points.

    Conclusions:

    • While reference point locations can vary significantly, corneal thickness at the apex and thinnest point is comparable.
    • The Orbscan II system allows flexibility in choosing central reference points without compromising corneal thickness measurement accuracy.
    • These findings support the interchangeable use of central reference points for corneal thickness evaluation in clinical practice.