Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Comparison of transcutaneous and endtidal CO2-monitoring for rigid bronchoscopy during high-frequency jet ventilation
M Simon1, R Gottschall, M Gugel
1Department of Anesthesiology and Intensive Care Medicine, University Hospital Jena, Germany. mark.simon@med.uni-jena.de
Objective:
To compare endtidal and transcutaneous respiratory monitoring of high-frequency jet ventilation (HFJV) in rigid bronchoscopy. Both techniques provide a noninvasive measurement of pCO2.
Methods:
High-frequency jet ventilation was applied via a rigid bronchoscope. Driving pressure (DP) was initially adapted to ensure normal ventilation. It was then changed twice by +/- 30% from the initial value. Endtidal and transcutaneous data were compared with arterial blood gas monitoring (ABG).
Results:
Results were analyzed separately for the time just after changing the driving pressure (10 min) and the remaining time until the next change of the driving pressure (4 min). The first part was called the dynamic phase, and the second part the steady-state phase. Correlation coefficient between endtidal capnography and ABG was 0.96 for the steady state and 0.94 for the dynamic phase, respectively. Bland-Altman analysis revealed a bias of -0.21 kPa with limits of agreement (LOA) 1.63 kPa for the steady state and -0.25 kPa, 2.08 kPa for the dynamic phase, respectively. Correlation coefficient between transcutaneous monitoring and ABG for the steady state phase was 0.83, and was 0.72 for the dynamic phase. Bland-Altman analysis resulted in a bias of -0.89 kPa with LOA - 3.84 kPa during steady state and 0.92 kPa, 4.06 kPa for the dynamic phase, respectively.
Conclusion:
Endtidal capnography offers accurate respiratory monitoring of HFJV. Transcutaneous monitoring showed a good correlation to ABG only during steady-state conditions. For the dynamic phase the accuracy was significantly lower. Thus, we cannot recommend transcutaneous respiratory monitoring for the specific indication of rigid bronchoscopy using HFJV.
More Related Videos
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Tracheostomy Suctioning II: Procedure
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)