Deterioration in the accuracy of the pulsair non-contact tonometer with use: need for regular calibration
P L Atkinson1, P K Wishart, J N James
1St Paul's Eye Hospital, Liverpool, UK.
Abstract:
The Pulsair non-contact tonometer (Keeler Pulsair: Keeler UK) has been shown to be a versatile instrument particularly suitable for screening for raised intraocular pressure. Although demonstrated to be accurate initially when compared to the Goldmann applanation tonometer no study has examined its long-term accuracy. Comparisons of three Pulsair tonometers with different degrees of usage with the Goldmann tonometer are described. Measurements were obtained from 64, 116, and 223 eyes in three separate comparative studies. Correlation coefficients of between 0.78 and 0.90 were obtained, the least used instrument being significantly more accurate than the two more extensively used instruments. Taking the Goldmann tonometer as the standard tonometer, and the aim to detect intraocular pressures of greater than 21 mmHg, sensitivities of 40%, 48%, and 85% for the three Pulsair tonometers respectively were shown. The Pulsair tonometer appears liable to a long-term drift in accuracy with use, and we suggest that provision is made for the regular re-calibration of the instrument.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Measurement of Blood Pressure
Distance Corrections
Errors in Taping