Objective:
Left ventricular assist devices have become an established method to bridge patients with end-stage cardiac failure to heart transplantation. Besides infection and bleeding, thromboembolism represents one of the most serious complications. We evaluated the value of microembolic signals in predicting thromboembolic events for individual patients and distinctive left ventricular assist device periods.
Methods:
Twenty patients (14 male) aged 23-57 years supported with the Novacor N100 left ventricular assist device were enrolled in this study. All patients were on effective anticoagulation, 12 patients additionally received antiplatelet therapy. Unilateral detection of microembolic signals was performed once weekly by insonation of the middle cerebral artery using transcranial Doppler sonography for 30 minutes duration. Evidence of clinically manifest thromboembolic events was based on regular questionnaires, clinical examinations, and results of diagnostic procedures.
Results:
During a cumulative follow-up of 3876 left ventricular assist device days, 44 thromboembolic complications occurred (incidence, 1.1%) in 15 out of 20 patients. A total of 360 transcranial Doppler sonography monitorings (range, 5-34 per patient) were performed with an overall microembolic signals prevalence of 35.3% and a microembolic signal mean of 2.3 +/- 9.2 per examination. There was a highly significant correlation between the individual microembolic signal activity and the respective incidence of clinical thromboembolism (r = 0.61-0.9; P <.01). Patients with additional antiplatelet treatment had significantly less thromboembolic complications (0.7%) and lower microembolic signal prevalence (18.3%) than those without (2.8% and 65.4%, respectively). Individual patients and left ventricular assist device months with clinical thromboembolization could be identified using the microembolic signal activity with moderate positive (0.37-0.7) and high negative predictive values (0.82-1.0).
Conclusions:
The amount of microembolic signals, serially detected in patients with the Novacor left ventricular assist device, is significantly associated with their incidence of embolic complications. The high negative predictive value of microembolic signals enables to identify those patients and left ventricular assist device periods with particularly low risk of clinical thromboembolization.
More Related Videos
07:25Predicting Amputation using Local Circulating Mononuclear Progenitor Cells in Angioplasty-treated Patients with Critical Limb Ischemia
Published on: September 22, 2020
10:28Interventional Diagnostic Procedure: A Practical Guide for the Assessment of Coronary Vascular Function
Published on: March 15, 2022
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Therapeutic Drug Monitoring: Affecting Factors
Acute Coronary Syndrome III: Diagnostic Studies
Venous Thrombosis I: Introduction
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis IV: Nursing Management