Comparison of the Retinomax autorefractor with hand-held retinoscopy in 1-year-old infants

Glen A Gole1, Philip J Schluter, Julie Hall

  • 1Community Child Health Service, Royal Children's Hospital, Brisbane, Queensland, Australia. g.gole@uq.edu.au

Insights

The Retinomax hand-held autorefractor shows good accuracy for infant vision screening, but cooperation is key. Results suggest it

Area of Science:

  • Ophthalmology
  • Pediatric Optometry
  • Medical Device Evaluation

Background:

  • Accurate refractive error assessment in infants is crucial for early detection of visual impairments.
  • Hand-held autorefractors offer potential for efficient screening in community settings.
  • Comparison of new autorefractor technology with established methods is needed.

Purpose of the Study:

  • To evaluate the accuracy and usability of the Retinomax hand-held autorefractor.
  • To compare Retinomax measurements with hand-held retinoscopy (HHR) in 1-year-old children.
  • To assess the impact of patient cooperation on measurement accuracy.

Main Methods:

  • Method comparison study involving 327 infants (46-81 weeks).
  • Refractive error measured using Retinomax and streak retinoscopy (HHR) under cycloplegia.
  • Sphere, cylinder, axis, and spherical equivalent data were recorded and compared.

Main Results:

  • Generally close agreement between Retinomax and HHR for refractive error measurements.
  • Retinomax readings were consistently ~0.3 D less hyperopic than HHR.
  • Significantly more girls struggled with the Retinomax examination than boys (P < 0.001).
  • Agreement between instruments decreased when patients struggled during examination (P < 0.001).

Conclusions:

  • The Retinomax appears to be a useful screening instrument for early childhood vision.
  • Patient cooperation significantly influences the accuracy of Retinomax measurements.
  • Careful consideration of usability and cooperation is necessary for adopting this screening tool.