D S Su1, H W Zandbergen, P C Tiemeijer
1Department of Inorganic Chemistry, Fritz-Haber-Institute of the Max-Planck-Society, D-14195 Berlin, Germany. dangsheng@fhi-berlin.mpg.de
Abstract:
Using single crystal V2O5 as a sample, we tested the performance of the new aberration corrected GATAN spectrometer on a monochromatised 200 kV FEG FEI (S)TEM. The obtained V L and O K ELNES were compared with that obtained in a common GATAN GIF and that in the new spectrometer, without monochromatised beam. The performance of the new instrumentation is impressive: recorded with an energy-resolution of 0.22 eV, the V L(3) edge reveals all the features due to the bulk electronic structure, that are also revealed in near-edge X-ray absorption fine structure (NEXAFS) with a much higher energy-resolution (0.08 eV). All features of the ELNES and NEXAFS are in line with a theoretical spectrum derived from band-structure calculations.
More Related Videos
Related Concept Videos
Atomic Emission Spectroscopy: Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Lab