On-line monitoring of a frontal chromatographic separation using inductively coupled plasma atomic emission spectroscopy
U Wenzel1, H Vijgen, W Ullrich
1Institut für Sicherheitsforschung und Reaktortechnologie, Forschungszentrum Jülich GmbH, 52425 Jülich, Germany. u.wenzel@fz-juelich.de
Abstract:
A test array is described employing a destructive analytical technique for the long-term monitoring of an industrial-scale separation process. As an example, we chose frontal chromatography as the separation and ICP/AES as the analytical method. The feed solution of the process was conveyed by a process pump via the separation unit to a sample station, where a small portion was diverted and transported by a roller pump into the spectrometer. We equipped our array with different loops for operating the process, calibrating the instrument and verifying the calibration. We obtained identical results for the different loops by absorbing the pulsation of the process pump and arranging for identical suction lines of the spectrometer pump. Based on the results, we redesigned the sample station for a technical application using only commercially available parts.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Gas Chromatography: Types of Detectors-II
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab