Echocardiographic considerations during deployment of the Helex Septal Occluder for closure of atrial septal defects
Related Experiment Video
Updated: Aug 3, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Leo Lopez1, Roque Ventura, Elizabeth M Welch
1Division of Pediatric Cardiology and the Heart Station, Miami Children's Hospital, Miami, Florida 33155, USA. llmd@llmd.net
Abstract:
The Helex Septal Occluder is a new device used to close atrial septal defects via interventional catheterization. In order to study the role of echocardiography during its use, and to describe the morphologic variants of defects suitable for closure with this occluder, we evaluated all patients undergoing intended closure of an atrial septal defect with the Helex occluder. A combination of transthoracic, transesophageal, three-dimensional, and intracardiac echocardiography were used before, during, and after the procedure to characterize anatomy, assess candidacy for closure, guide the device during its deployment, and evaluate results. Among the 60 candidates included in the study, 11 were excluded because of transesophageal echocardiographic and/or catheterization data obtained in the laboratory. Attempts at closure were successful in 46 patients, and unsuccessful in 3. We successfully treated four types of defects. These were defects positioned centrally within the oval fossa with appreciable rims along the entire circumference of the defect, defects with deficient or absent segments of the rim, defects with aneurysm of the primary atrial septum, and defects with multiple fenestrations. Follow-up transthoracic echocardiograms taken at a median of 7 months demonstrated no residual defects in 21, trivial residual defects in 17, and small residual defects in 8 patients. In 20 patients, three-dimensional reconstructions were used to characterize the morphology of the defect and the position of the device. Because transesophageal echocardiography was often limited by acoustic interference from the device, intracardiac echocardiography was utilized in 3 cases to overcome this limitation.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...