Related Experiment Videos
[Determination of trace Rh in organic residue by flame atomic absorption spectrophotometry]
Mei Han1, Fang Cheng, Shi-fang Gu
1Basic Experiment Center of Chemistry, Dalian University of Technology, Dalian 116024, China.
Guang Pu Xue Yu Guang Pu Fen Xi = Guang Pu
|August 28, 2003
Abstract:
The catalytic organic residue contained Rh was digested with HNO3 at 130 degrees C for 12 h. Trace noble metal Rh in catalytic organic residue was determined by flame atomic absorption spectrophotometry. Rh was atomized by air-acetylene flame at lamp Current of 7 mA. The methods of sample pretreatment for Rh in residue were compared in this paper. The recovery rates of this method were 95.3%-105.5% and the relative standard derivation was 0.9%. The method was applied to the analysis of some practical samples and the results obtained were satisfactory.