Related Experiment Videos
Design, construction and mechanical optimisation process of electrode with radial current flow in the scala tympani
P R Deman1, T M Kaiser, J J Dirckx
1Univ ENT Department, AZ St. Augustinus Hospital, Antwerp, Belgium.
Journal of Neuroscience Methods
|September 2, 2003
Abstract:
A 48 contact cochlear implant electrode has been constructed for electrical stimulation of the auditory nerve. The stimulating contacts of this electrode are organised in two layers: 31 contacts on the upper surface directed towards the habenula perforata and 17 contacts connected together as one longitudinal contact on the underside. The design of the electrode carrier aims to make radial current flow possible in the cochlea. The mechanical structure of the newly designed electrode was optimised to obtain maximal insertion depth. Electrode insertion tests were performed in a transparent acrylic model of the human cochlea.